The Priestley Smith Retinoscope operates on the principle of reflected light movement to determine the refractive error of the eye. It was developed by Ernest Priestley Smith, who contributed significantly to ophthalmic instrumentation in the late 19th century.
Principle:
The retinoscope uses a beam of light directed into the eye, which is then reflected back from the retina. The reflection, or retinal reflex, is observed through the retinoscope. The movement of this reflex changes depending on the refractive state of the eye, and the goal is to identify the point at which this reflex is neutralized (i.e., no movement is observed). This point corresponds to the correction required for the eye.
Key Concepts:
Far Point of the Eye:
The light reflected from the retina emerges from the eye as if it originates from the far point, which is the location where rays converge in a myopic eye or diverge in a hyperopic eye.
Neutralization:
The aim is to match the far point of the patient’s eye with the plane of the retinoscope. At this point, the retinal reflex is neutralized, appearing stationary.
With and Against Movements:
With Movement: Observed when the reflex moves in the same direction as the retinoscope. It indicates hyperopia or low myopia.
Against Movement: Observed when the reflex moves in the opposite direction. It indicates myopia.
Adding Lenses:
Convex or concave lenses are introduced in front of the patient’s eye to bring the far point to the plane of the retinoscope. The power of the lens required for neutralization determines the refractive error.
Design:
The Priestley Smith Retinoscope typically features:
A light source for illumination.
A rotating mirror to project the light beam.
An aperture through which the reflex is observed.
The retinoscopy technique using the Priestley Smith Retinoscope remains a fundamental method in clinical practice for objectively assessing refractive errors, laying the groundwork for modern retinoscopy and autorefractors.